Low temperature photoluminescence (PL) spectroscopy enables efficient investigation of the electronic structure of semiconductor materials and structures.
PL spectrometer in IPE consists of a Sumitomo optical He closed cycle cryostat coupled with THR 1000 monochromator (Jobin-Yvon) and FTIR spectrometer Nicolet 5700.
The PL setup enables sensitive and high resolution measurements in the following spectral (300-9000 nm) and temperature (3.5-300 K) ranges.
Grating spectrometer
FTIR spektrometer
The following light sources are available for excitation:
He-Ne laser, HNA 181, Zeiss Jena, excitation wavelength 632.8 nm
Ar ion laser Coherent Innova 300, excitation wavelength 514.5, 488 a 457.9 nm
He-Cd laser Kimmon IK3552R-G, excitation wavelength 325 nm